Personal exposure to static and time-varying magnetic fields during MRI system test procedures.

By F. de Vocht, F. Muller, H. Engels and H. Kromhout

Abstract

PURPOSE: To assess personal time-weighted average (TWA) static magnetic field and time-varying magnetic field exposure for several system testing tasks routinely conducted by engineers at an MRI manufacturing plant. MATERIALS AND METHODS: Personal exposure was measured using a personal dosimeter that measured B and dB/dt in all three orthogonal directions with a 1 Hz frequency. TWA exposure was calculated and random effects and linear mixed effects models were used to assess exposure levels and variability. RESULTS: Whereas full-body peak (2T) and TWA (200 mT/8 h) exposure limits for the static magnetic field originally proposed by ICNIRP were not exceeded, time-varying magnetic field peaks (dB/dt) did exceed the proposed threshold value (767.9 mT/s) suggested by IEEE on several occasions at the 3.0 Tesla (T) whole-body cylindrical system. CONCLUSION: TWA exposure levels are well below the proposed limit values, but peak exposure limits are exceeded during certain procedures performed on 1.0T to 3.0T systems

Year: 2009
OAI identifier: oai:dspace.library.uu.nl:1874/39560
Download PDF:
Sorry, we are unable to provide the full text but you may find it at the following location(s):
  • http://dspace.library.uu.nl:80... (external link)
  • Suggested articles