Article thumbnail
Location of Repository

Sorption Processes in Gas Sterilization in the Medical Sector

By A. Jordy and H. Suhr

Abstract

Sorption of ethylene oxide during and after gaseous sterilization is influenced by numerous factors. It was found that ethylene oxide desorption not only depends on material to be fumigated but also to a considerable degree on the wrapping material. Although polyethylene, polyamide (nylon), polytetrafluoroethylene (Teflon), silicone, aluminum, and glass beads contained no quantities of ethylene oxide detectable by gas chromatography after 72 h of aeration, residual amounts were definitely determined, even after 76 h of aeration in polypropylene, polystyrene, polyvinylchloride, paper products, and compound products of various plastics and paper mixtures. Desorption was, in all cases, found to be better when a mixture of ethylene oxide and methyl formate was used instead of pure ethylene oxide

Topics: Sterilization
Year: 1973
OAI identifier: oai:pubmedcentral.nih.gov:379856
Provided by: PubMed Central
Sorry, our data provider has not provided any external links therefore we are unable to provide a link to the full text.

Suggested articles


To submit an update or takedown request for this paper, please submit an Update/Correction/Removal Request.