oaioai:doaj.org/article:3f45286ee9624417be358cfe4d35a459

Invalidity of SUV Measurements of Lesions in Close Proximity to Hot Sources due to “Shine-Through” Effect on FDG PET-CT Interpretation

Abstract

It is well known that many technical and physiologic factors can affect the reliability of the standardized uptake value (SUV) on FDG PET-CT. Another potential problem of which we may be aware but has not been previously discussed is significant SUV overestimation of lesions in the direct neighborhood of large hot sources, namely, areas with high FDG uptake or activity such as a tumor, myocardium, urinary bladder, kidney, or gastrointestinal tract. The magnitude of SUV overestimation of the lesions directly neighboring the large hot sources is varied among the different cases, and it is possibly secondary to “shine-through” effect of the hot sources, which would warrant further systematic investigation such as phantom simulation experiment. If the lesion is in the close territory of the hot source, measured SUV is often overestimated and invalid. Visual interpretation should be used for evaluation of FDG avidity of the lesion

Similar works

Full text

thumbnail-image

Directory of Open Access Journals

Provided original full text link
oaioai:doaj.org/article:3f45286ee9624417be358cfe4d35a459Last time updated on 12/18/2014

This paper was published in Directory of Open Access Journals.

Having an issue?

Is data on this page outdated, violates copyrights or anything else? Report the problem now and we will take corresponding actions after reviewing your request.