Skip to main content
Article thumbnail
Location of Repository

Method Development and Validation for Simultaneous Estimation of Ketorolac and Sparfloxacin by RP-HPLC

By Rahul Singh, Ashish Pathak and Pooja Chawla*

Abstract

A modified simple, selective, rapid, precise reversed phase high performance liquid chromatography method has been developed and validated for the simultaneous estimation of ketorolac and sparfloxacin. The separation was made in a Hypersil-Keystone C-18 column using a methanol: water (60:40, v/v) (pH 3.1) as mobile phase at 308 nm. The mobile-phase flow rate and the sample volume injected were 0.9 ml/min and 20 µl, respectively. Retention time of sparfloxacin and ketorolac was found to be 3.181 and 4.473 minutes respectively. The correlation coefficient of both drugs was found to be 0.999. The accuracy of ketorolac was found to be 99.82% - 100.55% whereas for sparfloxacin, it was 99.76% - 99.89%. Over all % RSD was found to be less than 2%. The method was validated according to ICH guidelines with respect to linearity, accuracy, precision, robustness, specificity, etc. The developed method can be used for routine analysis of ketorolac and sparfloxacin in their pharmaceutical dosage forms

Topics: HPLC, Reverse phase, Ketorolac, sparfloxacin, Pharmacy and materia medica, RS1-441, Medicine, R, DOAJ:Pharmacy and materia medica, DOAJ:Medicine (General), DOAJ:Health Sciences
Publisher: Shri Manmohan Sharma Educational Society
Year: 2013
OAI identifier: oai:doaj.org/article:e6ab9cd056484764910e4b321be56f62
Journal:

Suggested articles


To submit an update or takedown request for this paper, please submit an Update/Correction/Removal Request.