The Pupillometric Precision of a Remote Video Eye Tracker

By Jeff Klingner

Abstract

To determine the accuracy and precision of pupil measurements made with the Tobii 1750 remote video eye tracker, we performed a formal metrological study with respect to a calibrated reference instrument, a medical pupillometer. We found that the eye tracker measures mean binocular pupil diameter with precision 0.10 mm and mean binocular pupil dilations with precision 0.15 mm

Topics: CR Categories, B.4.2 [Hardware, Input/Output and Data Communications—Input/Output Devices, I.4.9 [Computing Methodologies, Image Processing and Computer Vision—Applications, J.4 [Computer Applications, Social and Behavioral Sciences—Psychophysiology Keywords
Year: 2013
OAI identifier: oai:CiteSeerX.psu:10.1.1.359.9753
Provided by: CiteSeerX
Download PDF:
Sorry, we are unable to provide the full text but you may find it at the following location(s):
  • http://citeseerx.ist.psu.edu/v... (external link)
  • https://graphics.stanford.edu/... (external link)
  • Suggested articles


    To submit an update or takedown request for this paper, please submit an Update/Correction/Removal Request.