10.1016/S0735-1097(87)80323-6

Pitfalls in the diagnosis of periprosthetic valvular regurgitation by pulsed Doppler echocardiography

Abstract

Pulsed Doppler echocardiographic diagnosis of periprosthetic valvular insufficiency may be difficult. This report details the pulsed Doppler echocardiographic findings in two patients who developed severe periprosthetic mitral regurgitation after porcine mitral valve replacement. In both patients, mitral regurgitation was difficult to diagnose and left atrial turbulence, when detected, appeared localized, suggesting only mild mitral regurgitation. However, the combination of abnormally high peak transmitral diastolic flow velocity, with a normal pressure half-time, and increased flow velocity in the tricuspid regurgitant jet compatible with severe pulmonary hypertension, in the absence of other apparent left heart disease, suggested the correct diagnosis of severe mitral regurgitation in both cases. Techniques for optimal pulsed Doppler assessment of the mitral anulus region are emphasized, as are the theoretic advantages of continuous wave and color-coded pulsed Doppler echocardiography for detection of periprosthetic regurgitation

Similar works

Full text

thumbnail-image

Elsevier - Publisher Connector

Provided original full text link
Last time updated on 6/5/2019

This paper was published in Elsevier - Publisher Connector .

Having an issue?

Is data on this page outdated, violates copyrights or anything else? Report the problem now and we will take corresponding actions after reviewing your request.