Article thumbnail

A simple and accurate method for calibrating the oscillation amplitude of tuning-fork based AFM sensors

By Jinquan Liu, Andrea Callegari, Martin Stark and Majed Chergui

Abstract

We have developed a simple and accurate method for calibrating the amplitude of vibration of quartz tuning fork sensors commonly used in at. force- and near field optical-microscopy. Unlike interferometric methods, which require a complex optical setup, the method we present requires only a simple measurement of the electro-mech. properties of the tuning-fork oscillator and can be performed in a matter of minutes without disturbing the exptl. setup. Comparison with interferometric methods shows that an accuracy of better than few percent can be routinely achieved

Publisher: 'Elsevier BV'
Year: 2009
DOI identifier: 10.1016/j.ultramic.2008.08.009
OAI identifier: oai:infoscience.tind.io:134954
Download PDF:
Sorry, we are unable to provide the full text but you may find it at the following location(s):
  • http://infoscience.epfl.ch/rec... (external link)
  • Suggested articles