Background: To compare the intraocular pressure readings obtained with the iCare rebound tonometer and the 7CR non-contact tonometer with those measured by Goldmann applanation tonometry in treated glaucoma patients. Design: A prospective, cross sectional study was conducted in a private tertiary glaucoma clinic. Participants: 109 (54M:55F) patients including only eyes under medical treatment for glaucoma. Methods: Measurement by Goldmann applanation tonometry, iCare rebound tonometry and 7CR non-contact tonometry. Main Outcome Measures: Intraocular pressure. Results: There were strong correlations between the intraocular pressure measurements obtained with Goldmann and both the rebound and non-contact tonometers (Spearman r values ≥ 0.79, p 0.01) eyes. The corneal compensated measurements from the non-contact tonometer were significantly higher compared to the other tonometers (p ≤ 0.001). Conclusions: The iCare rebound tonometer and the 7CR non-contact tonometer measure IOP in fundamentally different ways to the Goldmann applanation tonometer. The resulting IOP values vary between the instruments and will need to be considered when comparing clinical versus home acquired measurements
Is data on this page outdated, violates copyrights or anything else? Report the problem now and we will take corresponding actions after reviewing your request.